Cookies on the OMEGA websites We use cookies on this website, these cookies are essential for the website to work correctly.If you continue without changing your settings, we'll assume that you are happy to receive all cookies on this website.To find out more information about these cookies please click here.
More information about how a load cell works in this link.
SPECIFICATIONS Excitation: 10 Vdc, 15 Vdc max Output: 3 mV/V ±0.25% 5-Point Calibration: (in Compression) 0%, 50%, 100%, 50%, 0% Linearity: ±0.03% FSO Hysteresis: ±0.02% FSO Repeatability: ±0.02% FSO Zero Balance: ±1% FSO Agency Approvals: FM Intrinsically Safe IS/I.II.III/1/CDEFG Operating Temp Range: -34 to 82°C (-30 to 180°F) Compensated Temp Range: 16 to 71°C (60 to 160°F) Thermal Effects: Zero: ±0.0018% FSO/°C Span: ±0.0018% FSO/°C Safe Overload: 150% of Capacity Ultimate Overload: 300% of Capacity Input Resistance: 360 Ohms minimum Output Resistance: 350 ±10 Ohms FS Deflection: 0.5 mm (0.02") typical Construction: 17-4 PH Stainless Steel Electrical: LCM501: 4.6 m (15 ft), 4-conductor shielded cable
LCM511: 6 pin DT02H-10-6P (or equiv.) Mating Connector LCM511: PT06F10-6S-R (not included)
† All amounts shown in EUR, GBP, or USD Note: Supplied with 5 point NIST traceable calibration certificate. Ordering Example: (1) LCM501-1K 0-1000 kgf Cantilever beam load cell with 4.6 m cable, £375.00